Enviro Testers Watch-Type Portable Radiation Dosimeter

Enviro Testers watch type personal dosimeter is mainly used for monitoring by X and gamma radiation.

SKU: ETERMD-108 Category:

Overview

Enviro Testers watch type personal dosimeter is mainly used for monitoring by X and gamma radiation. It caused by individual cumulative dose and dose equivalent rate, could stay wrist like a watch, is more suitable for interventional radiology treatment doses of radiation measurement and alarm, can display real-time data and alarm indication is given, and a data storage, alarm parameters setting, battery voltage detection, wireless charging and wireless data communications, and other functions. Alarm indication is given by Acousto-optic instructions. Its features are compact structure, waterproof and dustproof and can be systematically managed.

 

Technical Specifications

Measurement Scope Cumulative dose Hp(10): 0.0 mSv to 10 Sv; Dose rate Hp (10): 1 mSv/h to 500 mSv/h
Alarm Threshold Cumulative dose in 1 mSv to 10 Sv Adjustable between; Dose rate in 1 mSv/h to 500 mSv/h Adjustable between; The alarm time is adjustable between 99 hours and 59 minutes.
Alarm Method Sound and light alarm and LCD display indicator
Alarm Type Cumulative dose or dose rate exceeding threshold, battery under voltage, working overtime, detector failure, etc., with settable cumulative dose increment flash prompt
LCD Display Range Hp(10): 0.00 mSv to 9999 mSv, Hp(10): 0.01 mSv/h to 1000 mSv/h
Energy Response < ±30% (50 keV to 1.5 MeV)
Response Linearity <±20%  (10 mSv/h to 500 mSv/h)
Relative Inherent Error <±10%  (1 mSv/h, Cs-137)
Overload Characteristics >10 Sv/h
Power Source  Wireless charging mode, Built-in rechargeable lithium battery(4.2V,100mAh), Continuous working time is more than 10 days. Additional charging and wireless communication seat (as shown in the figure)
Under Voltage Indication The battery can still work for 8 hours after under voltage indication
Temperature Characteristics <±15% (14 ℉ to 104 ℉)(-10℃ to +40℃)
Humidity Characteristics <±10%(40% RH to 90% RH)
Data Storage Function When the instrument is turned off, the measurement data will be stored without power loss
Data Communication Data storage is transmitted to the computer wirelessly for system management
Feet Inches Φ 1.93 in x 0.63 in (Φ49mm × 16mm) (not including wristband)
Weight About 0.1 lb (0.045 Kg)