| Range |
pHmVTemperature |
0.00 pH to 14.00 pH1999.9 mV to 1999.9 mV32 °F to 212 °F (0.0 ℃ to 100.0 ℃) |
| Resolution |
pHmVTemperature |
0.01 pH0.1 mV32.18 °F (0.1 ℃) |
| Accuracy |
pHmVTemperature |
±0.01 pH±0.4 mV (-1500.0 to 1500.0); ± 1 mV ( -1999.9 to -1500.0 and 1500.0 to 1999.9)32.36 °F (±0.2 ℃) |
| pH Buffer Recognition |
US (pH 4.01, pH 7.00, pH 10.01) or NIST (pH 4.00, pH 6.86, pH 9.18), user selectable. |
| Buffer Temp range |
32 °F to 140 °F (0.0 ℃ to 60.0 ℃) |
| pH Calibration |
1-, 2- or 3-point calibration. |
| pH Electrode Offset Recognition |
±1.50 pH |
| pH Electrode Slope Recognition |
±30% at pH 4.00, pH 4.01, pH 9.18 or pH 10.01. |
| pH Temp Compensation |
Auto / manual 32 °F to 212 °F (0.0 ℃ to 100.0 ℃). |
| Input Impedance |
>1013 ohms. |
| Power |
6 X 1.5 V AAA batteries or 115 V/ 230 V AC adapter. |
| Operating Ambient Conditions |
From 32 °F to 122 °F (0 ℃ to 50 ℃), at 90% RH. |
| Case |
IP 54 waterproof |
| Dimensions |
5.9 in x 7.9 in x 2.8 in (150 mm x 203 mm x 72 mm) |
| Weight |
1.1 lb (504 g) |